H2 Hydrogen Sensor for low pressure carburizing

Measuring principle: Heat conductivity of measuring gas in nitrogen, ammonia, hydrogen and carbonic gases.
Technical Data ■ Measuring converter WLD / heat-conductive gas sensor ■ Measuring gas temperature: corresponding to process temperature ■ Connection temperature at KF flange: < 65°C ■ Measuring gas pressure: 3-30mbar, calibrated for 5 mbar abs. ■ Measuring components: Hydrogen: 10 - 100 Vol%, Carbonic gases ■ Measuring range : 0...100 Vol% ■ Output: 4-20 mA linear ■ Time response : Output delay 6...20s ■ Preheat time: Depending on sensor installation, up to 30 minutes at room temperature ■ Dimension: 110x70x240mm (WxHxD)
Accessories ■ Documentation ■ Power supply unit (optional) 24V/3A, short-circuit proof, output voltage 24V ± 10% ■ 2m connecting line with connected plug connector to sensor
Requirements for use ■ Convection pipe, depending on installation situation on request ■ Connection with KF16 vacuum flange
Example for mechanical connection of an H2-Sensor with convection pipe:

Data Sheet
H2-Sensor for low pressure carburizing
Language:

|